EU IVDR implementation date and the transition rules for IVDs

Time : Oct 02, 2026
Views:
EU IVDR implementation date: learn the 26 May 2022 requirements, legacy IVD transition deadlines, notified body milestones, and key compliance steps.

The EU IVDR implementation date was 26 May 2022. From that date, Regulation (EU) 2017/746 on in vitro diagnostic medical devices became fully applicable and replaced the former In Vitro Diagnostic Medical Devices Directive, Directive 98/79/EC (IVDD).

That date did not mean every IVD already supplied in Europe had to be immediately certified under IVDR. The regulation introduced transition arrangements for certain existing products—commonly called legacy devices—because the IVDR substantially expanded notified body involvement and changed the classification, evidence, and post-market obligations for many tests.

The practical question is therefore not simply whether an IVD was on the market before May 2022. It is whether the product qualifies for the transitional regime, what IVDR risk class it has, and whether the manufacturer meets the staged conditions needed to use the extended deadlines.

What changed on 26 May 2022?

Under the IVDD, only a limited group of higher-risk IVDs required notified body involvement before being placed on the EU market. Many manufacturers could self-declare conformity. The IVDR uses a risk-based classification system—Classes A, B, C, and D—and requires notified body assessment for most devices outside non-sterile Class A.

This is a material regulatory change. A test that was self-certified under the IVDD may be Class B or Class C under IVDR and therefore require a notified body certificate. The change affects not only the time required for certification, but also the depth of technical documentation, performance evaluation, quality management system review, post-market surveillance planning, and economic-operator controls.

The 26 May 2022 application date is thus the point at which new IVDs generally needed to comply with IVDR. It is not a universal expiry date for all IVDD-compliant devices, nor is it a guarantee that older documentation remains acceptable until the final transition deadline.

EU IVDR implementation date and the transition rules for IVDs

The current transition deadlines for legacy IVDs

The transition periods were extended by Regulation (EU) 2024/1860. The applicable end date depends on the device’s IVDR classification:

IVDR classification Latest transition end date Typical regulatory significance
Class D 31 December 2027 Highest-risk IVDs, including tests with significant implications for public health or transmissible agents
Class C 31 December 2028 High individual risk and/or moderate public-health risk devices
Class B and Class A sterile 31 December 2029 Moderate-risk devices and sterile Class A products requiring notified body involvement for sterility aspects

These dates are often described as “IVDR implementation deadlines,” but that wording can be misleading. They are not extra time to postpone all IVDR work. They are the outer limits for placing qualifying legacy devices on the market under a transitional route. Manufacturers must meet earlier milestones to remain eligible.

Non-sterile Class A devices are not included in these extended transition periods because they do not require notified body involvement under IVDR. Such products generally needed to comply with IVDR from 26 May 2022.

Which IVDs can use the transition period?

The transition is aimed at certain devices that were lawfully placed on the market under the IVDD and that require notified body involvement under IVDR. In broad terms, the device must be supported by an EC declaration of conformity drawn up before 26 May 2022 under the IVDD, and it must not have required notified body certification under the IVDD while requiring one under IVDR.

That distinction matters. A product cannot ordinarily be designed after the IVDR date and then introduced as an “old IVDD device” merely because it resembles an earlier product family. Likewise, a new product without a qualifying IVDD history does not gain access to the legacy-device deadlines.

Devices that already held an IVDD certificate issued by a notified body may be subject to separate provisions related to the validity of those certificates. Their position should be assessed against the specific certificate, its expiry status, the product scope, and the applicable provisions of Article 110 of the IVDR. The regulatory status of a portfolio should never be inferred solely from the device’s commercial launch date.

The conditions attached to the extended dates

A legacy device may continue under the transition route only if the manufacturer satisfies several conditions. The most important are not administrative formalities; they determine whether continued market access is legally defensible.

No significant change in design or intended purpose. The product must not undergo a significant change in its design or intended purpose after 26 May 2022. This is a frequent source of confusion. A manufacturer may need to maintain and improve a product, correct defects, manage suppliers, or update software. Some changes can be permissible, but a change that materially affects safety, performance, design, or intended purpose can remove the device from the transition route.

The Medical Device Coordination Group’s guidance on significant changes for legacy devices, MDCG 2022-6, is commonly used to evaluate this question. Its decision logic is useful because it distinguishes between routine quality-system activity and modifications that can alter the product’s regulatory identity. A change assessment should be documented rather than treated as a commercial or engineering decision alone.

Continued compliance with the IVDD. Until IVDR certification is obtained, the device must continue to comply with the IVDD requirements that applied to it. The transition does not suspend the older legal framework; it preserves a limited route under it.

No unacceptable risk. The device must not present an unacceptable risk to the health or safety of patients, users, or other persons, or to other aspects of public health protection. This places real weight on complaint handling, vigilance, corrective actions, performance monitoring, and distributor feedback. A legacy label does not shield a device from market surveillance or safety intervention.

IVDR quality management system. Manufacturers using the extended transition must have a quality management system compliant with Article 10(8) of the IVDR in place by 26 May 2025. This obligation is broader than preparing an application file. It concerns the system through which the manufacturer controls design, suppliers, production, risk, post-market activities, and regulatory compliance.

Notified body application and agreement milestones

The IVDR transition schedule also requires manufacturers to enter the notified body process early. The deadlines vary by device class:

  • For Class D devices, a formal application to a notified body must have been lodged by 26 May 2025, followed by a written agreement with that notified body by 26 September 2025.
  • For Class C devices, the application deadline is 26 May 2026, and the written agreement deadline is 26 September 2026.
  • For Class B and Class A sterile devices, the application deadline is 26 May 2027, and the written agreement deadline is 26 September 2027.

A written agreement is more than an expression of interest or an informal reservation with a notified body. It must cover the conformity assessment activity needed for the device. A manufacturer that waits until the final market-access date to begin selecting a notified body may find that it has already missed the relevant legal milestone.

The certification itself need not necessarily be completed by the application or agreement deadline. However, the manufacturer must remain within the transition conditions while the IVDR conformity assessment progresses. This creates a practical overlap period in which the legacy device remains on the market under the transition, while the IVDR technical file and quality-system assessment are being reviewed.

Why IVDR classification must be confirmed before planning a deadline

IVDR classification is the starting point for transition planning because it determines both the final date and the notified body milestones. The rules are set out in Annex VIII of the IVDR and are based on intended purpose and risk, not on the commercial value or physical complexity of the instrument.

For example, the same analytical technology can fall into different classes depending on what it detects, how the result is used, whether it is intended for screening or diagnosis, and what consequences an incorrect result may have. Tests for blood grouping, transmissible agents, cancer-associated markers, genetic information, or companion diagnostics can involve classification considerations that are not apparent from a product brochure.

Classification should also be performed for each relevant device, not only at platform level. An analyser, an assay, a calibrator, a control material, sample preparation component, or software may have different regulatory roles. A broad “laboratory system” description can obscure which items are separate devices and how each one should be classified.

Placing on the market is different from making available

Another important change introduced by Regulation (EU) 2024/1860 concerns the former “sell-off” concept. The IVDR no longer imposes a fixed sell-off deadline for devices lawfully placed on the EU market before the end of the applicable transition period.

In regulatory terms, placing on the market is the first making available of a specific device on the EU market. Making available refers to subsequent supply within the distribution chain, such as a distributor supplying stock to a laboratory, hospital, or another distributor.

The removal of the fixed sell-off deadline means that stock lawfully placed on the market before the relevant deadline may continue to be made available, subject to normal legal requirements and any applicable national measures. It does not mean that manufacturers can continue placing additional legacy devices on the market after their class-specific transition deadline.

This distinction matters for inventory planning. Importers and distributors should be able to demonstrate when stock was first placed on the EU market and preserve traceability records. Purchase orders, customs entries, invoices, warehouse records, and economic-operator documentation can become relevant where the status of inventory is questioned.

What the transition does not excuse

The transition periods were introduced to protect continuity of supply while manufacturers move into the new regulatory system. They do not create a lower standard of safety or a simplified version of IVDR compliance.

Manufacturers remain responsible for post-market surveillance, vigilance reporting, corrective actions, traceability, labeling controls, and cooperation with competent authorities. Importers and distributors also retain their own obligations to verify CE marking, documentation, labeling, storage and transport conditions, and the identity of the responsible economic operators.

For an IVD supplied from outside the EU, the EU authorized representative remains a central part of the compliance chain. The authorized representative’s name and address must appear as required, and the relationship should be supported by a written mandate. The presence of a distributor or commercial agent does not replace the need for an appropriately appointed authorized representative.

Laboratories should distinguish carefully between commercially supplied CE-marked IVDs and in-house devices developed and used within a health institution. In-house devices have a separate framework under Article 5(5) of the IVDR. They are not simply ordinary commercial devices that can be exempted from IVDR certification because they are used in a laboratory setting.

The practical compliance question is portfolio-specific

The EU IVDR implementation date is often searched as if it produced one answer for every diagnostic product. In practice, a reliable answer requires a device-by-device review. A portfolio may contain products already certified under IVDR, legacy devices with different class-specific deadlines, Class A products that should already be IVDR compliant, and products whose intended purpose changes make transition eligibility uncertain.

The most useful internal record is a regulatory status matrix that identifies the device and intended purpose, IVDR class, current conformity route, applicable transition date, notified body milestone, quality-system status, change-control history, and the evidence supporting each conclusion. This is more useful than a single company-level statement that the business is “IVDR ready.”

For supply-chain planning, the key date is not always the same as the commercial delivery date. A manufacturer may place a device on the market before the relevant deadline, while an importer, distributor, or laboratory receives it later. The documentation must show that the specific unit was lawfully placed on the market, not merely that the product model was eligible in principle.

IVDR timelines should therefore be treated as compliance boundaries tied to classification, documentation, notified body engagement, and controlled product change. The extended deadlines provide time, but only to manufacturers that actively progress toward IVDR conformity and preserve the conditions of the transitional route.

Have questions about the global trade trends discussed here, or planning your expansion and supply chain strategy? Leave a message below and let the trade experts at MTHH guide you.